An artificial intelligence-enhanced immunomagnetic bead platform for scalable melatonin quantification in heart failure

Xiaohan Liu ,  Zhu Chen ,  Zepeng Zhou ,  Fengling Chen ,  Yujuan Huang ,  Zhi Liu ,  Yajie Chen ,  Yuan Guo

Citation:  Xiaohan Liu, Zhu Chen, Zepeng Zhou, Fengling Chen, Yujuan Huang, Zhi Liu, Yajie Chen, Yuan Guo. An artificial intelligence-enhanced immunomagnetic bead platform for scalable melatonin quantification in heart failure[J]. Chinese Chemical Letters, 2027, 38(1): 112562. doi: 10.1016/j.cclet.2026.112562 shu

An artificial intelligence-enhanced immunomagnetic bead platform for scalable melatonin quantification in heart failure

English

  • Melatonin, an indoleamine synthesized by the pineal gland during the dark phase, is one of the most robust biomarkers of circadian phase [1]. Its secretion is tightly governed by light input to the suprachiasmatic nucleus, coordinating sleep-wake cycles, metabolic homeostasis, redox responses, and autonomic activity [2]. Disruptions in melatonin rhythms have been consistently associated with multiple chronic diseases, particularly heart failure (HF), in which reduced levels and dampened rhythmic amplitude correlate with oxidative injury, inflammation, and mitochondrial dysfunction [3-5]. Accordingly, accurate quantification of circulating melatonin may offer valuable insight into circadian status and disease risk in HF [6]. However, the low physiological concentration, prominent diurnal fluctuation, and stringent analytical requirements have made routine clinical melatonin quantification challenging.

    Several analytical platforms are currently used for melatonin measurement, including radioimmunoassay (RIA), enzyme-linked immunosorbent assay (ELISA), liquid chromatography-tandem mass spectrometry (LC-MS/MS), and chemiluminescent assays [7,8]. RIA provides high sensitivity but relies on radioactive reagents, posing regulatory and waste-disposal limitations [9]. ELISA is accessible and low-cost but constrained by antibody affinity, narrow dynamic range, enzymatic variability, and batch inconsistency [10,11]. LC-MS/MS represents the analytical gold standard with excellent specificity and sensitivity, yet requires expensive instrumentation, complex sample preparation, and technically demanding operation, limiting scalability [12]. Chemiluminescent methods offer increased throughput but depend heavily on reagent stability and closed analytical systems [13]. Collectively, existing approaches remain limited in sensitivity, scalability, cost-effectiveness, and clinical adaptability, underscoring the need for a more accessible and accurate analytical strategy.

    Melatonin presents inherent biochemical challenges for detection. As a small, hydrophobic indole derivative (232.28 Da) with low polarity, it easily associates with proteins and lipids within biological fluids [14]. These properties complicate immunogen design, as carrier-coupled conjugation may distort native epitopes and reduce antibody specificity. Moreover, the picomolar physiological concentration demands high analytical sensitivity [15]. Despite these challenges, immunoassays remain conceptually attractive owing to their specificity, operational simplicity, and compatibility with limited sample volumes, provided that their inherent limitations can be effectively addressed [16]. Conventional competitive immunoassays, however, often suffer from compressed signal ranges, limited linearity, suboptimal antibody stability, and batch variability, restricting their quantitative accuracy [17]. These limitations highlight the necessity for an optimized immunochemical approach that enables sensitive, accurate, and cost-effective quantification of melatonin.

    Immunomagnetic bead platforms provide several advantages for small-molecule assays [18]. Magnetic beads offer a high surface-area-to-volume ratio, tunable surface chemistry, and efficient mixing kinetics that improve antigen-antibody interactions, particularly under low-abundance conditions [19,20]. Their compatibility with flow cytometry enables single-particle resolution, wide dynamic ranges, and multi-parametric fluorescence readouts. In competitive immunoassays, the high-density antibody presentation on beads improves signal discrimination and enhances quantification precision [21]. Despite challenges related to fluorescence stability, bead aggregation, and batch-to-batch variability, immunomagnetic systems present a compelling foundation for next-generation melatonin assays. Importantly, the same system attributes that confer analytical advantages also generate intrinsically multidimensional and heterogeneous signal outputs, increasing analytical complexity and limiting the effectiveness of conventional calibration strategies, thereby necessitating the development of more advanced quantitative frameworks [22].

    Artificial intelligence (AI)-based algorithms have emerged as powerful tools for interpreting complex analytical signals and enabling ultrasensitive quantification of low-abundance targets [23,24]. In bead-based sensing platforms, the multidimensional fluorescence and scatter data generated by flow cytometry present nonlinear relationships and interdependent features that are difficult to resolve using traditional single-parameter calibration models [22,25]. Extreme gradient boosting (XGBoost), a tree-based ensemble learning algorithm, has shown favorable performance in modeling nonlinear signals, resisting noise, maintaining interpretability, and accommodating limited sample sizes, attributes that are critical for quantitative biomarker assessment [26,27]. By exploiting multidimensional feature integration rather than isolated signal parameters, XGBoost can effectively suppress background interference and enhance concentration prediction accuracy, making it well suited for integration with immunomagnetic bead-based detection systems.

    To address the limitations of existing melatonin assays, we developed an AI-enhanced immunomagnetic bead method that integrates competitive fluorescent immunorecognition with XGBoost-based quantitative modeling. This platform enables sensitive, precise, low-cost, and scalable melatonin quantification with an expanded dynamic range and simplified workflow. Benchmarking against ELISA confirmed comparable analytical performance, while application to serum from HF patients demonstrated feasibility and robustness in clinical samples. This approach provides a practical solution for routine melatonin assessment and may facilitate broader translational use of circadian biomarkers across cardiovascular research and beyond.

    To implement this analytical strategy, we began by verifying the successful and stable covalent attachment of phycoerythrin (PE)-conjugated anti-melatonin antibodies to carboxylated magnetic beads (MBs@Ab) (Fig. 1A). By laser confocal microscopy revealing a distinct fluorescent halo surrounding the antibody-conjugated beads, a feature was absent in their unconjugated counterparts (Fig. 1B). This successful immobilization was further substantiated by flow cytometry, which demonstrated a significant shift to higher PE fluorescence intensity in the MBs@Ab population compared to control beads (Figs. 1C and D).

    Figure 1

    Figure 1.  Immobilization of antibody against melatonin on magnetic beads. (A) Diagram of antibody labeling and magnetic bead conjugation. (B) Images of magnetic beads pre/post-conjugation. Scale bar: 10 µm. (C, D) Flow cytometry of magneti beads pre/post-conjugation. (E) Absorption spectra of antibody solutions at graded concentrations. (F) Regression of the absorbance at 280 nm and the concentration of antibody solution. (G) The quantity of antibody consumed. (H) PE positive rate after the conjugation (the numbers denote experimental batches). (I, J) PE positive rate of MBs@Ab maintained in the dark (I: 4 ℃; J: 4 ℃ and room temperature (RT)). Data are presented as mean ± standard deviation (SD) (n = 3).

    Beyond the initial validation, the efficiency and uniformity of antibody conjugation were systematically quantified. A standard curve derived from antibody absorbance at 280 nm (Figs. 1E and F) was used to determine the residual antibody concentration in the post-conjugation supernatant. The results showed that 61.1% of the input antibodies were successfully conjugated to the surface of the magnetic beads. This value likely reflects the deliberate use of excess antibody during the coupling process rather than the intrinsic coupling efficiency (Fig. 1G) [28,29].

    To further characterize the MBs@Ab population, the entire bead cohort was evaluated by flow cytometry. A fluorescence threshold (PE intensity > 1000 a.u.) was defined to discriminate conjugated from unconjugated beads, revealing that >99% of MBs@Ab exhibited PE fluorescence (Fig. 1H). These findings demonstrate that the conjugation procedure achieved highly uniform surface modification, with nearly all magnetic beads effectively functionalized by antibody attachment.

    The stability of the antibody-bead conjugates was assessed due to its relevance for assay implementation and reproducibility. PE-positivity of MBs@Ab remained above 99.5% over a 30-day monitoring period when stored at 4 ℃ in the dark (Fig. 1I). Comparable stability was maintained at room temperature, with PE-positivity exceeding 99.5% after 24 h (Fig. 1J). These findings indicate that the conjugation process generates a highly uniform and stable bead population suitable for storage and downstream analytical use.

    Then, we synthesized a fluorescein-labeled melatonin analog (FITC-MT) to serve as a competitive tracer and defined reaction conditions that balance signal robustness with quantitative range (Fig. 2A). After incubation with FITC-MT and MBs@Ab, a clear FITC ring was observed around the beads, confirming specific binding of FITC-MT by the immobilized antibody (Fig. 2B). Flow cytometry further established a FITC-positive threshold (>10,000 a.u.) and demonstrated a marked shift only when beads were incubated with FITC-MT, not with 5-MT alone (Fig. S1A in Supporting information), indicating that unmodified reactants do not contribute to the FITC signal on beads.

    Figure 2

    Figure 2.  Validation of FITC-MT and establishment of the reaction system. (A) Diagram of FITC-MT synthesis. (B) Images of co-incubation of FITC-MT with MBs@Ab. Scale bar: 10 µm. (C) FITC positive rate of varying molar ratios of 5-MT reacted with FITC–NHS. (D, E) FITC positive rate of MBs@Ab with varying concentrations of FITC-MT (incubation time: 0.5-h for D and 3-h for E). (F) FITC positive rates of MBs@Ab with FITC-MT (0.5 and 1.0 nmol/mL) for different durations. (G) FITC positivity rates of MBs@Ab with 300 pg/mL melatonin and FITC-MT (0.25, 0.5, and 0.75 nmol/mL). Data are presented as mean ± SD (n = 3).

    To evaluate product quality and determine the influence of reaction stoichiometry, the molar ratio of 5-MT to FITC–N-hydroxysuccinimide (NHS) was systematically varied during synthesis (Fig. S1B in Supporting information). When FITC–NHS was present at equimolar or higher levels relative to 5-MT, >99% of beads surpassed the FITC fluorescence threshold, indicating efficient conjugation and sufficient tracer purity (Fig. 2C). These findings confirm that FITC-MT is a specific and sufficiently pure tracer suitable for subsequent assay development.

    Optimization studies were next performed to define tracer concentration and incubation duration that would enable near-equilibrium binding while maintaining adequate competitive capacity. Across a titration series, both the proportion of FITC-positive beads and mean fluorescence intensity increased with rising tracer concentration and, to a lesser extent, with incubation times ranging from 0.5 h to 3 h (Figs. 2D and E). Extending incubation to 12 h resulted in only marginal gain in FITC-positivity at 0.5 nmol/mL and no additional increase at 1.0 nmol/mL (Fig. 2F, Fig. S1C in Supporting information). Given that native melatonin exhibits higher binding affinity for the antibody than FITC-MT, a 3-h incubation period was selected to achieve sufficient equilibration for competitive assays while avoiding unnecessary procedural delay.

    A working tracer concentration was subsequently defined based on competitive performance within the intended analytical range. When 300 pg/mL melatonin was introduced into reactions containing 0.25, 0.5, and 0.75 nmol/mL FITC-MT, the corresponding FITC-positive bead fractions were 32.5%, 63.2%, and 99.7% (Fig. 2G). To ensure robust fluorescence discrimination across 0–300 pg/mL and prevent premature saturation that would compress the dynamic range, 0.25 nmol/mL FITC-MT was selected as the optimal concentration for quantitative assays.

    Together, these experiments validate FITC-MT as a specific tracer and establish the working conditions for the competitive reaction, which involve 3 h of incubation with 0.25 nmol/mL FITC-MT. With the competitive conditions established, we next evaluated the analytical performance of the assay using standard melatonin solutions.

    Using the optimized competitive conditions, as shown in Fig. S2 (Supporting information), analytical performance was evaluated across a set of serially diluted melatonin calibrators (Fig. S2A). Representative flow-cytometry profiles derived from these calibrators are shown in Fig. S2B. Linear regression analyses based on the proportion of FITC positive beads, median fluorescence intensity, and mean fluorescence intensity yielded coefficients of determination of 0.3882, 0.4551, and 0.7515, respectively (Figs. S2C–E). These results indicate that single-parameter linear regression provides limited accuracy and insufficient quantitative performance.

    To clarify the basis of this limitation, bead-level heterogeneity was examined across key workflow stages. Forward- and side-scatter profiles showed progressively greater dispersion from unconjugated beads to MBs@Ab, with further amplification within the full reaction matrix (Fig. S2F). The fraction of events within the homogeneous cluster declined from 95.9% (unconjugated) to 65.0% (MBs@Ab) and to 9.27% under reaction conditions (Fig. S2F). Confocal microscopy confirmed increased size variability and micro-aggregation in the reaction environment (Fig. S2G). Consistently, forward scatter-side scatter (FSC-SSC) heatmaps revealed marked distributional shifts and the emergence of distinct bead subpopulations across stages (Fig. S2H).

    Collectively, these findings illustrate that single-parameter fluorescence metrics and linear calibration models fail to account for the substantial bead-to-bead heterogeneity, aggregation-associated signal distortion, and inherent nonlinearities introduced during the assay. These characteristics necessitate a multiparametric analytical framework capable of leveraging the full cytometric feature space and modeling the complex, heterogeneous relationships between signal patterns and melatonin concentration.

    We therefore developed an AI-enhanced pipeline to infer melatonin concentrations directly from multiparametric flow-cytometry data (Fig. 3A). Bead-level events first underwent standardized preprocessing, including gating, spectral compensation, quality control, and intensity normalization. Following normalization, bead-level measurements were extracted into feature vectors comprising FSC, SSC, allophycocyanin (APC), FITC, and PE across all pulse types. Subsequently, feature engineering techniques were applied to identify an optimal subset of informative variables for model development, following the removal of collinear features (Fig. 3B).

    Figure 3

    Figure 3.  The flow cytometry data of magnetic beads analyzed by XGBoost. (A) Diagram of the model processing of melatonin standard sample flow cytometry data. (B) The features correlation heat map of the model (S1-S3, cluster coding). (C, D) RMSE and R2 on the training and validation sets across training rounds; the dashed line marks the early-stopping round. (E) SHAP feature importance bar plot.

    Model training was performed using XGBoost, with training and validation sets allocated according to a predefined split ratio. Early stopping was implemented to prevent overfitting. As illustrated by the learning curves, increasing training iterations resulted in progressive root mean square error (RMSE) reduction on both the training and validation sets, reaching a minimum at the early stopping point, while R2 increased and plateaued (Figs. 3C and D). The model showed high consistency between the training set (R2 = 0.9964 ± 0.0023, RMSE = 8.7023 ± 2.3475) and the validation set (R2 = 0.9836 ± 0.0086, RMSE = 8.6655 ± 2.3267). The small performance gap confirms that the model is robust, and no overfitting was observed.

    To enhance model interpretability and quantify the contribution of individual features, SHapley additive exPlanations (SHAP) analysis was performed. The results identified S3_FITC-W, S2_FITC-W, and S3_Ratio as the most influential predictors, consistent with the central role of FITC-derived signals in melatonin detection (Fig. 3E). Additional contributions arose from features associated with population proportions, FSC reflecting bead size, SSC indicative of bead structural properties, and PE fluorescence related to antibody loading. These observations also support the hypothesis that magnetic bead uniformity is altered in complex biological matrices.

    Using the same preprocessing and feature engineering pipeline, conventional four- and five-parameter logistic models based on the single most important feature, as well as multivariate regression, least absolute shrinkage and selection operator (LASSO) regression, and random forest models, were independently evaluated for comparison (Fig. S3 in Supporting information). Relative to traditional single-parameter linear calibration, multiparametric approaches improved the monotonic relationship between signal patterns and melatonin concentration and reduced noise arising from intra- and inter-batch bead heterogeneity.

    To enable practical use, we developed a standalone software tool that automates the full workflow (Fig. S4A in Supporting information). This platform is designed to directly ingest raw flow-cytometry files, execute the trained XGBoost model, and generate final melatonin concentrations without requiring manual feature engineering, calibration procedures, or specialized computational expertise.

    The operational workflow is optimized for scalable use (Fig. S4B in Supporting information). Users load raw data files corresponding to calibration standards and test samples, after which the software autonomously performs all preprocessing steps, including gating, normalization, and feature extraction. The resulting feature vectors are then processed through the embedded XGBoost model to yield calculated melatonin concentrations. Outputs are displayed within the graphical user interface, and all results can be exported as structured tabular files to support downstream documentation and statistical analysis. By integrating preprocessing, model inference, and result reporting into a unified interface, the platform reduces manual intervention, ensures analytic consistency across batches, and enables rapid, scalable deployment of the AI-enhanced bead assay (AI-bead) in translational and clinical research environments.

    To further assess the generalizability and practical performance of the AI-bead, we quantified melatonin concentrations in both standard solutions and real serum samples. The results indicated that, within the calibration range, the AI-bead assay exhibited good agreement with the expected concentrations (Fig. 4A and Table S1 in Supporting information). When evaluating endogenous melatonin in serum samples, measurements obtained from the AI-bead demonstrated good consistency with results generated by the commercial ELISA (Fig. 4B). These findings indicate that the AI-bead demonstrates acceptable generalizability and provides quantification accuracy potentially exceeding that of existing ELISA platforms, with performance expected to further improve as additional data are incorporated into model refinement.

    Figure 4

    Figure 4.  Application of AI-bead in heart failure. (A) Scatter plot of standard concentration and detection by AI-bead. (B) Bland-Altman plots of serum melatonin by ELISA and AI-bead. (C) Serum melatonin levels in HF patients and healthy controls. (D-H) The relationship between serum melatonin and age, logarithmic B-type natriuretic peptide (Log BNP), left ventricular ejection fraction (LVEF), left ventricular end-diastolic diameter (LVEDD), and aortic diameter. *P < 0.05.

    The assay was next deployed to assess serum melatonin profiles in an HF cohort compared with healthy controls. Baseline clinical characteristics are summarized in Table S2 (Supporting information). Mean serum melatonin was significantly lower in individuals with HF (109.2 ± 9.7 pg/mL) relative to healthy participants (136.5 ± 6.6 pg/mL, P < 0.05, Fig. 4C), a pattern consistent with the circadian dysregulation reported in HF populations.

    To evaluate clinical correlates, associations between melatonin concentrations and key clinical indices were examined. Melatonin levels were inversely associated with age (Pearson correlation coefficient: −0.326, P < 0.05), whereas no significant correlations were observed with B-type natriuretic peptide, left ventricular ejection fraction, left ventricular end-diastolic diameter, or aortic diameter (Figs. 4D-H). These relationships are aligned with prior evidence suggesting that melatonin more closely reflects systemic circadian and vascular milieu rather than acute hemodynamic stress [30]. Collectively, the strong analytical agreement with ELISA, coupled with biologically coherent clinical associations, underscores the reliability and translational utility of the AI-bead assay for serum melatonin quantification in HF.

    In this study, we developed an AI-enabled immunomagnetic bead assay for quantitative melatonin detection. The platform combines competitive immunorecognition on antibody-conjugated magnetic beads with flow-cytometric acquisition of multidimensional features and XGBoost-based concentration modeling. This strategy achieved analytical performance comparable to ELISA. The optimized model was further integrated into a graphical interface to streamline calibration and reporting within the flow-cytometry workflow. Application to serum from patients with HF demonstrated reliable detection of circulating melatonin, supporting the feasibility of this approach for translational circadian research.

    Conventional melatonin assays, including RIA, ELISA, LC-MS/MS, and chemiluminescent immunoassays, each carry inherent methodological constraints [7]. Immunoassays often suffer from batch-dependent variability and limited dynamic range, whereas LC-MS/MS, although analytically robust, requires expensive instrumentation, complex sample preparation, and is not readily scalable for large-cohort or longitudinal circadian studies [7,31]. These limitations have collectively hindered the widespread clinical and research applications of melatonin measurements, particularly in contexts where temporal resolution and cost-effectiveness are essential [32]. Our assay was designed explicitly to address these unmet needs.

    Compared with ELISA kits based on similar solid-phase immunoassay principles, the improved performance of the assay arises from the combined advantages of immunomagnetic beads and data-driven analysis. The high surface-area-to-volume ratio of magnetic beads enables dense and stable antibody immobilization [20,33], thereby enhancing assay sensitivity, an essential requirement for detecting melatonin at its low physiological concentrations. Moreover, the suspension-based reaction format promotes efficient antigen–antibody interactions in small-volume samples, facilitating effective capture of native melatonin [10]. Beyond antibody affinity alone, quantitative performance in competitive immunoassays is governed by complex binding and dissociation kinetics, incubation dynamics, and non-equilibrium processes [34]. However, conventional analytical approaches based on single-parameter readouts or predefined calibration models are often insufficient to resolve subtle variations arising from processes and are susceptible to noise, inter-batch variability, and bead heterogeneity, ultimately limiting quantitative accuracy and reproducibility [22].

    To overcome these limitations, this study demonstrates a shift from single-parameter calibration toward multiparametric, data-driven quantification of low-abundance circadian biomarkers. By integrating fluorescence intensity, bead size, and structural features derived from flow-cytometry data, the XGBoost model effectively accommodates nonlinear relationships and mitigates the influence of noise and aberrant bead populations. This multiparametric framework captures a broader spectrum of assay variability associated with binding dynamics and bead heterogeneity, enabling more stable and reproducible melatonin quantification. Rather than relying on static calibration curves [26], the model facilitates consistent identification of informative bead populations and improves robustness across experimental conditions, highlighting the analytical advantage of algorithm-assisted interpretation for complex immunoassay data. The strategy of applying algorithmic models to leverage multiple parameters for the quantification of low-abundance biomarkers may offer a useful complementary analytical approach for the detection of other trace-level analytes.

    The broadened detection range of this platform is clinically meaningful. Melatonin circulates at extremely low physiological concentrations, exhibits pronounced circadian oscillation, and may rise substantially following exogenous supplementation-particularly given the widespread use of high-dose over-the-counter melatonin products [1,32]. Most existing assays often fail to cover both physiologic nadirs and supraphysiologic peaks [7]. In contrast, the immunomagnetic bead system, through training on expanded datasets, enables reliable melatonin detection across a wide concentration range, from low to high concentrations. This feature positions the assay as a valuable tool for both circadian biology research and the monitoring of melatonin supplementation.

    We further demonstrated the clinical applicability of the platform by measuring serum melatonin in HF patients, a condition characterized by circadian disruption, metabolic dysregulation, and autonomic imbalance [35]. Consistent with prior reports, patients with HF exhibited reduced melatonin levels with clinically coherent associations, supporting the biological plausibility of circadian impairment in this population [36]. These findings indicate that the platform can detect clinically meaningful differences in low-abundance circadian biomarkers and may be extendable to other disorders associated with circadian misalignment, including diabetes, hypertension, and sleep-related conditions [30,37,38].

    This platform also offers potential for large-scale circadian phenotyping. Once the XGBoost model is trained and baseline calibration established for each bead batch, subsequent samples can be analyzed without reconstruction of standard curves, enabling high-throughput implementation. This efficiency, combined with low assay cost, could facilitate population-level circadian studies previously limited by technical and financial barriers. Collectively, these findings support the immunomagnetic bead, AI platform as a sensitive, scalable, and cost-efficient method for melatonin quantification with potential integration into cardiovascular research, precision therapeutics, and longitudinal monitoring.

    Several limitations warrant consideration. Although the platform was validated in patients with HF, further evaluation across diverse disease states, populations, and laboratory settings is necessary to confirm robustness and inter-site reproducibility. Importantly, the present method is not intended to replace LC-MS/MS, which remains the gold standard for absolute melatonin quantification, but rather to serve as a complementary analytical tool optimized for accessibility, throughput, and routine application. Additional comparative studies with LC-MS/MS, particularly across higher concentration ranges, will be valuable for defining the practical scope and optimal use cases of this approach.

    Taken together, this study addresses the long-standing challenge of reliably quantifying melatonin in clinical settings by developing an AI-enhanced immunomagnetic bead assay that integrates the high-efficiency capture characteristics of magnetic beads with the multiparametric modeling strengths of the XGBoost algorithm. This platform provides high analytical accuracy, low cost, and scalability compared with existing methods. Validation in a HF cohort revealed significantly reduced circulating melatonin levels consistent with prior pathophysiological evidence, underscoring the assay’s applicability for evaluating circadian disruption in clinical populations. Given its robustness, operational simplicity and quantitative precision, this method provides a practical tool for circadian phenotyping in diverse chronic diseases, population-based research, pharmaceutical testing and routine health assessment. Overall, this AI-enhanced immunomagnetic assay represents a scalable and translationally relevant platform for advancing circadian biomarker measurement in both research and clinical practice.

    This study adheres to the Declaration of Helsinki (World Medical Association, 2013), and was approved by the Ethics Committee of Zhuzhou Hospital Affiliated to Xiangya School of Medicine, Central South University (approval No. KY2024079–01). All patients were informed and provided consent prior to inclusion in the study.

    Xiaohan Liu: Writing – original draft, Methodology, Investigation, Formal analysis, Data curation. Zhu Chen: Writing – review & editing, Supervision, Investigation, Conceptualization. Zepeng Zhou: Formal analysis, Data curation. Fengling Chen: Formal analysis, Data curation. Yujuan Huang: Formal analysis, Data curation. Zhi Liu: Visualization, Formal analysis. Yajie Chen: Visualization, Formal analysis. Yuan Guo: Writing – review & editing, Supervision, Investigation, Funding acquisition, Conceptualization.

    The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

    This work was supported by The National Natural Science Foundation of China (No. 82372581), and Health Research Project of Hunan Provincial Health Commission (No. W20242007).

    Supplementary material associated with this article can be found, in the online version, at doi:10.1016/j.cclet.2026.112562.


    1. [1]

      M. Jia, A. Luo, M. Wei, et al., Sleep Med. Rev. 82 (2025) 102123. doi: 10.1016/j.smrv.2025.102123

    2. [2]

      T. Lei, H. Hua, H. Du, et al., Arch. Toxicol. 98 (2024) 395–408. doi: 10.1007/s00204-023-03647-5

    3. [3]

      Y. Guo, Z. He, Z. Chen, et al., Theranostics 13 (2023) 1826–1842. doi: 10.7150/thno.82069

    4. [4]

      M. Xin, F. Bi, C. Wang, et al., J. Adv. Res. 69 (2025) 495–514. doi: 10.1016/j.jare.2024.04.005

    5. [5]

      S. Crnko, B.C. Du Pre, J.P.G. Sluijter, et al., Nat. Rev. Cardiol. 16 (2019) 437–447. doi: 10.1038/s41569-019-0167-4

    6. [6]

      F. Fagiani, D. Di Marino, A. Romagnoli, et al., Signal Transduct. Target. Ther. 7 (2022) 41. doi: 10.1038/s41392-022-00899-y

    7. [7]

      D.J. Kennaway, J. Pineal Res. 67 (2019) e12572. doi: 10.1111/jpi.12572

    8. [8]

      X. Xiao, Z. Huang, G. Luo, et al., Trends Analyt. Chem. 195 (2026) 118616. doi: 10.1016/j.trac.2025.118616

    9. [9]

      A.W. Hsing, T.E. Meyer, S. Niwa, et al., Cancer Epidemiol. Biomark. Prev. 19 (2010) 932–937. doi: 10.1158/1055-9965.EPI-10-0004

    10. [10]

      D.J. Kennaway, Chronobiol. Int. 38 (2021) 616–619. doi: 10.1080/07420528.2021.1886112

    11. [11]

      G. Fatima, V.P. Sharma, N.S. Verma, Spinal Cord 54 (2016) 364–367. doi: 10.1038/sc.2015.176

    12. [12]

      A. Munoz-Jurado, D. López, B.M. Escribano, Food Biosci. 59 (2024) 104222. doi: 10.1016/j.fbio.2024.104222

    13. [13]

      K.K.A.C. Monteiro, M.E. Shiroma, L.L. Damous, et al., Antioxidants 13 (2024) 439. doi: 10.3390/antiox13040439

    14. [14]

      D.J. Kennaway, J. Pineal Res. 69 (2020) e12657. doi: 10.1111/jpi.12657

    15. [15]

      C.D. Flynn, D. Chang, A. Mahmud, et al., Nat. Rev. Bioeng. 1 (2023) 560–575. doi: 10.1038/s44222-023-00067-z

    16. [16]

      R.J. Tran, K.L. Sly, J.C. Conboy, Anal. Chem. 92 (2020) 13163–13171. doi: 10.1021/acs.analchem.0c02286

    17. [17]

      S. Kim, H.J. Kang, J.M. Kim, et al., Anal. Chim. Acta 1378 (2025) 344697. doi: 10.1016/j.aca.2025.344697

    18. [18]

      H. Park, M.P. Hwang, K.H. Lee, Int. J. Nanomed. 8 (2013) 4543–4552.

    19. [19]

      E. Zhang, J. Cao, J. Cheng, et al., Chin. Chem. Lett. 37 (2026) 111109. doi: 10.1016/j.cclet.2025.111109

    20. [20]

      Y. Li, S. Liu, Y. Wang, et al., Biosensors 13 (2023) 903. doi: 10.3390/bios13100903

    21. [21]

      A. Adan, G. Alizada, Y. Kiraz, et al., Crit. Rev. Biotechnol. 37 (2017) 163–176. doi: 10.3109/07388551.2015.1128876

    22. [22]

      S.X. Leong, Y.X. Leong, C.S.L. Koh, et al., Chem. Sci. 13 (2022) 11009–11029. doi: 10.1039/d2sc02981b

    23. [23]

      B.A. Taha, A.J. Addie, L.F.Z. Kolie, et al., Chin. Chem. Lett. 37 (2026) 111955. doi: 10.1016/j.cclet.2025.111955

    24. [24]

      P. Mukherjee, S. Sen, A. Das, et al., Biosens. Bioelectron. 271 (2025) 117023. doi: 10.1016/j.bios.2024.117023

    25. [25]

      L.A. Clarke, V. Shah, F. Arrigoni, et al., J. Thromb. Haemost. 6 (2008) 1025–1032. doi: 10.1111/j.1538-7836.2008.02953.x

    26. [26]

      T. Chen, P. Liu, Y. Ge, et al., Chem. Eng. J. 506 (2025) 160080. doi: 10.1016/j.cej.2025.160080

    27. [27]

      J. Si, Y. Bao, F. Chen, et al., Eur. Heart J. Digit. Health 6 (2025) 82–95. doi: 10.1093/ehjdh/ztae092

    28. [28]

      P. Carl, Ii Ramos, M.A. Segundo, et al., PLoS One 14 (2019) e0218686. doi: 10.1371/journal.pone.0218686

    29. [29]

      P. Yu, B. Chen, B. Xie, et al., Chin. Chem. Lett. 37 (2026) 111695. doi: 10.1016/j.cclet.2025.111695

    30. [30]

      G. Cornelissen, Y. Watanabe, J. Siegelova, et al., Biol. Rhythm Res. 50 (2019) 46–62. doi: 10.1080/09291016.2018.1491193

    31. [31]

      X. Huang, G. Mazza, Crit. Rev. Food Sci. Nutr. 51 (2011) 269–284. doi: 10.1080/10408398.2010.529193

    32. [32]

      F. Nduhirabandi, G.J. Maarman, Molecules 23 (2018) 1819. doi: 10.3390/molecules23071819

    33. [33]

      G. Yang, Z. Li, R. Usman, et al., Chin. Chem. Lett. 36 (2025) 109930. doi: 10.1016/j.cclet.2024.109930

    34. [34]

      X. Wang, L. Cohen, J. Wang, et al., J. Am. Chem. Soc. 140 (2018) 18132–18139. doi: 10.1021/jacs.8b11185

    35. [35]

      S. Mia, R. Sonkar, L. Williams, et al., JACC Basic Transl. Sci. 8 (2023) 1141–1156. doi: 10.1016/j.jacbts.2023.03.016

    36. [36]

      T. Misaka, A. Yoshihisa, T. Yokokawa, et al., J. Pineal Res. 66 (2019) e12564. doi: 10.1111/jpi.12564

    37. [37]

      M.T. Smith, C.S. Mccrae, J. Cheung, et al., J. Clin. Sleep Med. 14 (2018) 1231–1237. doi: 10.5664/jcsm.7230

    38. [38]

      S. Vacas, E. Mcinrue, M.A. Gropper, et al., Anesth. Analg. 123 (2016) 206–212.

  • Figure 1  Immobilization of antibody against melatonin on magnetic beads. (A) Diagram of antibody labeling and magnetic bead conjugation. (B) Images of magnetic beads pre/post-conjugation. Scale bar: 10 µm. (C, D) Flow cytometry of magneti beads pre/post-conjugation. (E) Absorption spectra of antibody solutions at graded concentrations. (F) Regression of the absorbance at 280 nm and the concentration of antibody solution. (G) The quantity of antibody consumed. (H) PE positive rate after the conjugation (the numbers denote experimental batches). (I, J) PE positive rate of MBs@Ab maintained in the dark (I: 4 ℃; J: 4 ℃ and room temperature (RT)). Data are presented as mean ± standard deviation (SD) (n = 3).

    Figure 2  Validation of FITC-MT and establishment of the reaction system. (A) Diagram of FITC-MT synthesis. (B) Images of co-incubation of FITC-MT with MBs@Ab. Scale bar: 10 µm. (C) FITC positive rate of varying molar ratios of 5-MT reacted with FITC–NHS. (D, E) FITC positive rate of MBs@Ab with varying concentrations of FITC-MT (incubation time: 0.5-h for D and 3-h for E). (F) FITC positive rates of MBs@Ab with FITC-MT (0.5 and 1.0 nmol/mL) for different durations. (G) FITC positivity rates of MBs@Ab with 300 pg/mL melatonin and FITC-MT (0.25, 0.5, and 0.75 nmol/mL). Data are presented as mean ± SD (n = 3).

    Figure 3  The flow cytometry data of magnetic beads analyzed by XGBoost. (A) Diagram of the model processing of melatonin standard sample flow cytometry data. (B) The features correlation heat map of the model (S1-S3, cluster coding). (C, D) RMSE and R2 on the training and validation sets across training rounds; the dashed line marks the early-stopping round. (E) SHAP feature importance bar plot.

    Figure 4  Application of AI-bead in heart failure. (A) Scatter plot of standard concentration and detection by AI-bead. (B) Bland-Altman plots of serum melatonin by ELISA and AI-bead. (C) Serum melatonin levels in HF patients and healthy controls. (D-H) The relationship between serum melatonin and age, logarithmic B-type natriuretic peptide (Log BNP), left ventricular ejection fraction (LVEF), left ventricular end-diastolic diameter (LVEDD), and aortic diameter. *P < 0.05.