Electrochemical Sensor Based on Ni12P5 Nanoparticles for Sensitive Determination of Glucose

Wen XU Xun ZHOU Jin-Ming XU Han XU Fei-Fei TAO

Citation:  XU Wen, ZHOU Xun, XU Jin-Ming, XU Han, TAO Fei-Fei. Electrochemical Sensor Based on Ni12P5 Nanoparticles for Sensitive Determination of Glucose[J]. Chinese Journal of Inorganic Chemistry, 2019, 35(4): 729-736. doi: 10.11862/CJIC.2019.073 shu

基于Ni12P5纳米粒子的电化学传感器用于灵敏测定葡萄糖

    通讯作者: 陶菲菲, feifeitao@usx.edu.cn
  • 基金项目:

    安徽省教育厅自然科学研究项目 KJHS2017B06

    安徽省教育厅自然科学研究项目 KJ2018A0409

    浙江省科技厅项目 2017C33209

    安徽省教育厅自然科学研究项目(No.KJHS2017B06, KJ2018A0409)和浙江省科技厅项目(No.2017C33209)资助

摘要: 通过改进的热溶剂胶体合成法制备了单分散的Ni12P5纳米粒子,并利用X射线衍射、透射电子显微镜、X射线光电子能谱、X射线能谱对Ni12P5纳米粒子的晶体结构、化学组成和形貌等进行了表征。基于单分散Ni12P5纳米粒子研制出的非酶葡萄糖传感器具有出色的性能,其快速响应时间小于3 s,检测范围广(0.002~4.2 mmol·L-1),灵敏度高达1 572 mA·L·mol-1·cm-2,检测限低至0.8 μmol·L-1。此外,该传感器在用于人体血液中葡萄糖的实际检测中取得了满意的效果。

English

  • Diabetes is one of the top ten diseases that would cause disability and even death in the world. Glucose concentration in human blood is one of the most key markers for the diagnoses and management of diabetes mellitus. Therefore, it is of paramount importance for developing a simple, sensitive and selective method to assay for glucose. Currently, a myriad of techniques have been utilized to determine glucose, including high performance liquid chromatography (HPLC)[1], chemiluminescence[2], fluorescence[3] and electro-chemistry[4-7]. Among these methods, the approach of electrochemical-based detection can allow a high-sensitive in situ detection with short response time, wide linear range and low cost. Currently, various glucose electrochemical biosensors have been developed and they are typically classified into enzyme-based and non-enzyme-based sensors. In 1962, Clark and Lyons reported the first enzymatic electrochemical biosensor for the determination of glucose[8]. Since then, three generations of glucose electrochemical sensors have been developed; they are all fabricated based on the glucose oxidase (GOx) enzymes. The enzyme-based sensor is highly sensitive and selective, but the natural enzymes are usually expensive and difficult to be fixed for the wide applications due to the typical drawbacks including time barrier, hydrophobic energy barrier, size barrier, etc. Especially, the sensor is easily affected by environmental temperature and pH value[9], and the enzyme fixed in the sensor tends to lose its corres-ponding enzymatic activity in the varied performing systems. Therefore, it is still a challenge to explore effective non-enzyme based glucose electrochemical biosensor to reduce and overcome the intrinsic shortcomings in a facile strategy.

    For integrating the non-enzyme based biosensors, electrochemical approach has been certified as an effective way. For examples, noble metals and alloys, transition metal chalcogenides and nanostructured transition metal chalcogenides/carbon composites have been fabricated as electrode materials to detect glucose effectively[10-14]. Recently, transition metal phosphides, as a sort of important functional materials, which have attracted wildly great interests in various catalysis reactions for hydro-desulfurization[15], oxygen reduc-tion[16], hydrogen and oxygen evolution via water splitting[17-20], due to their superior and instinct electrical conductivity[21]. However, transition metal phosphides have been rarely reported as electrode materials for non-enzyme based glucose detection except for the ones of Ni2P, CoP and NiCoP[22-24].

    Herein, it is presented that Ni12P5 NPs were synthesized via a modified one-pot hot-solution colloidal preparation method. Then, the obtained Ni12P5 NPs which have been constructed to shape a sensitive non-enzyme based glucose sensor exhibited high electrocatalytic activity to glucose oxidation, with the virtues of a quick response time less than 3 s, a broad linear concentration ranging from 0.002 to 4.2 mmol·L-1, a high sensitivity up to 1 572 mA·L·mol-1·cm-2, and a detection limit as low as 0.8 μmol·L-1. The performances were favorably comparable to the most nickel-based catalysts, as seen the literature listed in below discussion. The results demonstrate that the electrode based on Ni12P5 NPs can be applied as a glucose sensor with the advantages of high sensitivity, short response time and reproducibility. Additionally, it can be used to detect the glucose in the human blood serum with satisfactory result.

    Nickel(Ⅱ) acetylacetonate hydrate (Ni(acac)2·xH2O, 95%) was bought from Tokyo Chemical Industry(TCI). Oleylamine (OAm, 80%~90%), tri-n-octylphosphine (TOP, 90%) and 1-octadecene (ODE, 90%) were provided by Alfa Aesar. D-(+)-Glucose, dopamine (DA), ascorbic acid (AA), uric acid (UA) and Nafion solution (5%(w/w)) were provided by Sigma-Aldrich. Lactose (Lac), fructose (Fru), sodium hydroxide, ethanol and toluene were supplied by Sinopharm Chemical Reagent Ltd. All reagents were used directly as purchased.

    The synthesis of the Ni12P5 NPs was adopted from Wang′s work with some modifications[25]. Briefly, 0.128 g Ni(acac)2, 1 mL TOP, 3 mL OAm and 2 mL ODE were introduced into a 100 mL three-necked flask and continuously magnetically stirred at room temperature under argon flow. Next, the mixed solution was heated to 140 ℃ and held for 30 min to remove impurities such as dissolved oxygen and low boiling point solvent in the solution, and then the device was heated to 270 ℃ at a rate of 10 ℃·min-1 and held for one hour. Finally, the apparatus was naturally cooled to normal temperature, then the black product was centrifuged five times with a mixed solution of toluene and ethanol and dried at 50 ℃ under vacuum for further use.

    The bare glassy carbon electrode (GCE) was polished to the mirror with an Al2O3 emulsion having a particle size of 0.3 and 0.05 μm, and the electrode was ultrasonically cleaned with water, dilute nitric acid (VH2O:VHNO3=1:1, 65%(w/w) HNO3) and ethanol each for 1 min. 4.0 mg of the prepared Ni12P5 powder and 30 μL Nafion solution were added to 1.0 mL ethanol-water solution (VH2O:VEtOH=1:1), and ultrasonically dispersed for one hour to form a well homogeneous suspension. Then 5.0 μL of the suspension was dropped onto the surface of GCE and dried in air overnight.

    The Philips Xpert PRO X-ray diffractometer (Cu , λ=0.154 178 nm, 40 kV, 50 mA, 2θ=20°~80°) was used to get the crystalline structure and phase purity of the as-synthesized Ni12P5 particles. Trans-mission electron microscopy (TEM) images reflecting the shape and size were obtained on a Hitachi H-7650 TEM (100 kV). X-ray photoelectron spectroscopy (XPS) were acquired on a Thermo ESCALAB 250 spectro-meter (Al , 15 kV, 10 mA) to determine the chemical compositions and the valence of the Ni12P5 NPs. The energy dispersive spectroscopy (EDS) was characterized by scanning electron microscope (SEM, Hitachi S3400N, 15 kV).

    Cyclic voltammetry (CV) and chronoamperometry (CA) were carried out on an electrochemical workstation (CHI660E, ChenHua) using a typical three-electrode system, the Ni12P5 modified electrode (φ=3 mm) served as the working electrode, a platinum sheet (1 cm2) as an auxiliary electrode and a saturated Ag/AgCl as the reference electrode. Every electrochemical measure-ment was carried out under normal temperature in 0.1 mol·L-1 NaOH solution (pH=13). The water used in experiments was deionized water (18.2 MΩ·cm).

    The crystal structure of the as-prepared Ni12P5 NPs was confirmed by XRD, as represented in Fig. 1a. The diffraction peaks located at 32.73°, 35.81°, 38.41°, 41.76°, 44.42°, 46.96°, 48.96°, 54.04°, 56.16°, 60.14°, 68.59° and 74.85° observed in the pattern could be indexed to (310), (301), (112), (400), (330), (240), (312), (510), (501), (242), (161) and (352) planes of tetragonal Ni12P5 (PDF No.74-1381)[26], respectively. There were not any obvious peaks from impurities and deviates, indicating the high purity of the samples.

    Figure 1

    Figure 1.  (a) XRD patterns of the Ni12P5 NPs; (b) Survey XPS spectrum for Ni12P5 NPs; High-resolution XPS spectra of Ni2p (c) and P2p regions (d)

    To reveal the chemical components and electronic state of the Ni12P5 NPs, the samples were measured by XPS. As shown in Fig. 1b, Ni, P, C and O were present in the product without other impurity. Fig. 1c exhibits the XPS spectrum for Ni12P5 in the Ni2p. The split peaks at 870 and 873.6 eV accompanied by one satellite peak at 880.4 eV were found in the Ni2p1/2 window[27], while the Ni2p3/2 region showed three peaks at 852.7, 855.7 and 860.9 eV[28]. The peaks at 852.7 and 870 eV were attributed to Ni in Ni12P5[29, 30], while those at 855.7 and 873.6 eV fitted well with oxidized Ni species[27, 30]. The peak at 852.7 eV was characteristics of Ni metal (852.5~852.9 eV), indicating that the Ni in Ni12P5 has a very small positive charge[24]. The P2p XPS spectrum (Fig. 1d) exhibited a doublet at 130.35 and 129.55 eV, corresponding to P2p1/2 and P2p3/2 respectively[30-31] and the peak at 133.2 eV in good agreement with oxidized phosphorus species formed on the surface of Ni12P5 because of air contact[32]. The binding energy of P2p3/2 (129.55 eV) was very close to that of zero-valent P (130.0 eV)[28, 33], suggesting that the P in Ni12P5 has a very small negative charge. The peaks of O and C could be assigned to the presence of oxygen, water and carbon dioxide molecules absorbed by the surface as well as hydrocarbons from the XPS machine itself[34].

    Fig. 2a is a representative low-magnified TEM image of the as-synthesized Ni12P5 NPs, it is clearly found that the Ni12P5 NPs were uniform with monodisperse size. In detail, the size of the monodisperse nanoparticles was determined to be about 9 nm in diameter based on the high-magnification transmission electron microscope image, as seen in Fig. 2b. Meanwhile, the chemical composition of the Ni12P5 NPs was further measured by energy dispersive X-ray spectroscopy (EDX), as shown in Fig.S1. The result reveals that the nanoparticles were composed of 68.07%(n/n) Ni and 31.93%(n/n) P with an atomic ratio of approximately 2.13:1 for nickel to phosphorus. This value approaches the stoichiometric ratio of Ni to P in Ni12P5.

    Figure 2

    Figure 2.  Typical TEM images of Ni12P5 NPs with low magnification (a) and high magnification (b)

    To evaluate the electrocatalytic activity of the Ni12P5 NPs to glucose oxidation, we dropped the Ni12P5 NPs on the surface of GCE to form the modified electrode performed in 0.1 mol·L-1 NaOH solution. In Fig. 3a, there was a pair of redox peaks at 0.606 and 0.429 V observed for the Ni12P5 NPs in absence of glucose, which could be connected with the Ni(Ⅲ)/Ni(Ⅱ) redox couples[22], corfirmed by XPS survey spectrum after CV treatment (Fig.S2~Fig.S4). When adding glucose, the peak potential of glucose oxidation was very close to that of Ni(Ⅱ) oxidation to Ni(Ⅲ)[13], and the adsorption of glucose and the oxidized intermediates on the active sites of the Ni12P5 based electrode, so the anodic peak potential shifted a little in the positive direction[35], but an obvious increase of anodic current density appeared, which shows that there is an electrocatalytic activity of Ni12P5 to glucose oxidation. In contrast, in case of GCE without addition of Ni12P5 (inset), there was just a featureless voltammetric current within the potential range of interest observed no matter whether glucose was introduced or not. The results confirm that the electrocatalytic activity to glucose oxidation is benefited from the employment of the Ni12P5 NPs. Meanwhile, it is noted that the current response of the Ni12P5/GCE electrode was very sensitive when adding glucose in 0.1 mol·L-1 NaOH solution, as seen in Fig. 3b. It further demonstrates that the Ni12P5/GCE is highly efficient for electrooxidation of glucose. In basic media, anodic scanning can lead to the formation of NiOx/Ni(OH)x on the surface of the Ni12P5 NPs[20, 30], so it can be concluded the mechanism of the electrocatalysts for Ni12P5 as the following equations, equivalent to the one reported in literature[36]:

    Figure 3

    Figure 3.  (a) CVs of Ni12P5/GCE (1, 2) and bare GCE (3, 4) in 0.1 mol·L-1 NaOH with the presence (1, 3) and absence (2, 4) of 0.1 mmol·L-1 glucose at a scan rate of 50 mV·s-1; (b) Amperometric response of the Ni12P5/GCE (1) and bare GCE (2) at 0.6 V with successive adding glucose of the same concentration in 0.1 mol·L-1 NaOH solution

    ${\rm{Ni}}{\left( {{\rm{OH}}} \right)_2} + {\rm{O}}{{\rm{H}}^ - } \to {\rm{NiO}}\left( {{\rm{OH}}} \right) + {{\rm{H}}_{\rm{2}}}{\rm{O + }}{{\rm{e}}^ - } $

    (1)

    ${\rm{NiO}}\left( {{\rm{OH}}} \right) + {\rm{glucose}} \to {\rm{Ni}}{\left( {{\rm{OH}}} \right)_2} + {\rm{glucolactone}} $

    (2)

    The effect of scan rate on electrooxidation of glucose was carried by CV with 0.1 mmol·L-1 glucose in 0.1 mol·L-1 NaOH solution (Fig. 4a). When the scan rate increasing from 10 to 150 mV·s-1, the oxidation peak positively shifted, whereas the reduction peak negatively shifted, both the oxidation and reduction peak current density increased continuously, and the current density of oxidation peak showed a good linear relationship with the square root of the scan rate (R2=0.995 8), as shown in Fig. 4b. The result clearly indicates that the electrooxidation of glucose on the Ni12P5 modified electrode belongs to a diffusion-controlled process[34].

    Figure 4

    Figure 4.  (a) CV curves of Ni12P5/GCE in 0.1 mol·L-1 NaOH with 0.1 mmol·L-1 glucose at various scan rates; (b) Linear relationship of oxidation current density vs the square root of scan rate

    Fig. 5a presents the electrochemical response of the electrode in 0.1 mol·L-1 NaOH solution at 0.6 V with the continuous injection of different concentrations of glucose. The modified electrode showed a quick response to the varying concentration of glucose. When adding the glucose, the oxidation current increased immediately and reached the steady-state current in less than 3 s, the current density was linear with the glucose concentration over a wide range of 0.002 to 4.2 mmol·L-1 (R2=0.998 9) and the sensitivity was 1 572 mA·L·mol-1·cm-2. The limit of detection was calculated as 0.8 μmol·L-1 (signal-to-noise ratio S/N=3), with the results shown in Fig. 5b. These values were comparable to most electrochemical glucose sensors constructed on Ni based nanomaterials, as listed in Table 1.

    Figure 5

    Figure 5.  (a) Amperometric response of the Ni12P5/GCE recorded at 0.60 V in 0.1 mol·L-1 NaOH solution with consecutive addition of various concentrations of glucose; (b) Linear fitting curve between current density and glucose concentrations

    Table 1

    Table 1.  Comparison between Ni12P5/GCE and other Ni-based electrochemical sensors for the detection of glucose
    下载: 导出CSV
    Electrode material Sensitivity/
    (mA·L·mol-1·cm-2)
    Linear range/
    (mmol·L-1)
    Detection limit/
    (μmol·L-1)
    Reference
    Ni7S6/GCE 271.8 0.005~3.7 0.15 [37]
    NiO-MWCNT/CPEa 6 527 0.001~14 19 [38]
    NiO HHs-Nafion/GCE 1 052.8 0.5~4.5 1.2 [39]
    Ni(OH)2/rGO 11 400 0.02~30 15 [40]
    NiS 7.43 0.005~0.045 0.32 [41]
    Ni(OH)2/NCFb 1 342.2 0.005~9.15 0.8 [42]
    Ni2P/Ti 2 289.5 0.001~6.0 0.4 [22]
    c-Ni(OH)2 HRc 1 569 0.002~3.8 0.6 [43]
    Ni-NPs/TiO2NTs 700.2 0.004~4.8 2 [36]
    Ni12P5/GCE 1 572 0.002~4.2 0.8 This work
    aCPE: carbon paste electrode; bNCF: nitrogen-doped carbon foam; cHR: hollow nanorod

    Selectivity is one of the most important parameters for evaluating the performance of sensor. Besides glucose, there are other sugars such as Fru or Lac and active substance such as DA, AA and UA in human blood. The normal value of glucose in the blood (4~7 mmol·L-1) is at least ten times higher than other interferent species (< 0.1 mmol·L-1)[44]. The experiment was carried out at the Ni12P5 modified electrode by adding 1.0 mmol·L-1 glucose in 0.1 mol·L-1 NaOH with 0.1 mmol·L-1 AA, UA, DA, Fru and Lac. As exhibited in Fig. 6, the current density increased greatly when adding glucose, whereas the other interferents showed weak current responses. This result suggests that the determination of glucose by our approach is not interfered by the other interferents in the blood, and non-enzymatic glucose biosensor developed on this method exhibits high anti-interference ability.

    Figure 6

    Figure 6.  (a) Amperometric response of the Ni12P5/GCE with consecutive additions of 1 mmol·L-1 glucose and 0.1 mmol·L-1 DA, UA, AA, Fru in 0.1 mol·L-1 NaOH solution at 0.60 V; (b) Variation in the response current density of Ni12P5/GCE toward 0.1 mmol·L-1 glucose in 0.1 mol·L-1 NaOH for 30 days

    To measure the reproducibility, 0.1 mmol·L-1 glucose was detected under the same conditions using five electrodes fabricated in the same manner, which produced a relative standard deviation (RSD) of 4.97%, showing a high reproducibility of our biosensor. The long-term stability of the sensor based on Ni12P5 NPs was determined every five days by CA in 0.1 mol·L-1 NaOH solution, the prepared sensor was kept in laboratory atmosphere when not in use. As depicted in Fig. 6b, the current density did not show a sharp drop and 91.6% of the initial current response was retained after 30 days. These results demonstrate that the developed sensor is stable with good repeatability and reproducibility.

    To examine the practicality of the biosensor, we applied the Ni12P5/GCE to detect the concentration of glucose in human blood serum obtained from the local hospital. 30 μL human blood serum was directly diluted with 0.1 mol·L-1 NaOH solution, then the amperometric test was performed at an operating potential of 0.6 V and glucose concentration was calculated by linear equation, with results shown in Table 2. The results obtained from our sensor are in accordance with those tested by an automatic biochemical analyzer, and the recovery percentage is close to 100%. It should be confirmed that the developed biosensor can be applied for testing glucose with sufficient accuracy.

    Table 2

    Table 2.  Test results of glucose in human blood serum (n=3)
    下载: 导出CSV
    Sample Concentration of glucose/(mmol·L-1) Recovery/%
    Hospital Our sensor Added Measured value
    1 4.15 4.23 5 9.27 100.8
    2 4.96 5.02 5 10.15 102.6
    3 4.36 4.45 5 9.38 98.6

    In summary, a sensitive nonenzymatic electro-chemical sensor based on the Ni12P5 NPs was designed for the first time to detect glucose. Owing to the high conductivity and excellent catalytic properties, the biosensor based Ni12P5 NPs displayed excellent electrocatalytic activity towards oxidation of glucose in alkaline medium, high stability, low detection limit and wide linear range were obtained in the experi-ment. Especially, the biosensor could greatly eliminate the influence of interfering substances such as AA, DA and UA based on the intensive investigations. Thus, the sensor based on the Ni12P5 NPs is promising for the routine detection of glucose.

    Supporting information is available at http://www.wjhxxb.cn

    1. [1]

      Diana D D, Carlos-Eduardo N C, Luz-Patricia R S, et al. J. Chromatogr. B, 2015, 975:18-23 doi: 10.1016/j.jchromb.2014.10.039

    2. [2]

      Zhong X, Chai Y Q, Yuan R. Talanta, 2014, 128:9-14 doi: 10.1016/j.talanta.2014.03.071

    3. [3]

      Shan X Y, Chai L J, Ma J J, et al. Analyst, 2014, 139:2322-2325 doi: 10.1039/C3AN02222F

    4. [4]

      Sun Y L, Yang H Y, Yu X H, et al. RSC Adv., 2015, 5:70387-70394 doi: 10.1039/C5RA13383A

    5. [5]

      Ahmad R, Tripathy N, Hahn Y B, et al. Dalton Trans., 2015, 44:12488-12492 doi: 10.1039/C5DT01664A

    6. [6]

      Lv J, Kong C C, Xu Y, et al. Sens. Actuators B, 2017, 248:630-638 doi: 10.1016/j.snb.2017.04.052

    7. [7]

      Mondal S, Madhuri R, Sharma P K. J. Mater. Chem. C, 2017, 5:6497-6505 doi: 10.1039/C7TC01411B

    8. [8]

      Clark L C, Lyons C. Ann. N.Y. Acad. Sci., 1962, 102:29-45

    9. [9]

      Ensafi A A, Ahmadi N, Rezaei B. Sens. Actuators B, 2017, 239:807-815 doi: 10.1016/j.snb.2016.08.088

    10. [10]

      Ensafi A A, Zandi-Atashbar N, Rezaei B, et al. RSC Adv., 2016, 6:60926-60932 doi: 10.1039/C6RA10698F

    11. [11]

      Saraf M, Natarajana K, Mobin S M. Dalton Trans., 2016, 45:5833-5840 doi: 10.1039/C6DT00670A

    12. [12]

      Li H, Guo C Y, Xu C L. Biosens. Bioelectron., 2015, 63:339-346 doi: 10.1016/j.bios.2014.07.061

    13. [13]

      Sonkar P K, Ganesan V, Abraham John S, et al. RSC Adv., 2016, 6:107094-107103 doi: 10.1039/C6RA16064F

    14. [14]

      Hui N, Wang W T, Xu G Y, et al. J. Mater. Chem. B, 2015, 3:556-561 doi: 10.1039/C4TB01831A

    15. [15]

      Prins R, Bussell M E. Catal. Lett., 2012, 142:1413-1436 doi: 10.1007/s10562-012-0929-7

    16. [16]

      Chen K Y, Huang X B, Wan C Y, et al. RSC Adv., 2015, 5:92893-92898 doi: 10.1039/C5RA21385A

    17. [17]

      Tian J Q, Liu Q, Cheng N Y, et al. Angew. Chem. Int. Ed., 2014, 53:9577-9581 doi: 10.1002/anie.201403842

    18. [18]

      You B, Jiang N, Sheng M L, et al. ACS Catal., 2016, 6:714-721 doi: 10.1021/acscatal.5b02193

    19. [19]

      Stern L A, Feng L G, Song F, et al. Energy Environ. Sci., 2015, 8:2347-2351 doi: 10.1039/C5EE01155H

    20. [20]

      Wang C D, Jiang J, Ding T, et al. Adv. Mater. Interfaces, 2016, 3:1500454 doi: 10.1002/admi.201500454

    21. [21]

      Carenco S, Portehault D, Boissiere C, et al. Chem. Rev., 2013, 113:7981-8065 doi: 10.1021/cr400020d

    22. [22]

      Chen T, Liu D N, Lu W B, et al. Anal. Chem., 2016, 88:7885-7889 doi: 10.1021/acs.analchem.6b02216

    23. [23]

      Sun Q Q, Wang M, Bao S J, et al. Analyst, 2016, 141:256-260 doi: 10.1039/C5AN01928A

    24. [24]

      Wang Z, Cao X Q, Liu D N, et al. Chem. Commun., 2016, 52:14438-14441 doi: 10.1039/C6CC08078B

    25. [25]

      Wang C D, Ding T, Sun Y, et al. Nanoscale, 2015, 7:19241-19249 doi: 10.1039/C5NR05432J

    26. [26]

      Zhang H J, Feng Y Y, Zhang Y, et al. ChemSusChem, 2014, 7:2000-2006 doi: 10.1002/cssc.v7.7

    27. [27]

      Pan Y, Liu Y Q, Liu C G. J. Power Sources, 2015, 285:169-177 doi: 10.1016/j.jpowsour.2015.03.097

    28. [28]

      Huang Z P, Chen Z B, Chen Z Z, et al. ACS Nano, 2014, 8:8121-8129 doi: 10.1021/nn5022204

    29. [29]

      Pan Y, Yang N, Chen Y J, et al. J. Power Sources, 2015, 297:45-52 doi: 10.1016/j.jpowsour.2015.07.077

    30. [30]

      Han A L, Chen H L, Sun Z J, et al. Chem. Commun., 2015, 51:11626-11629 doi: 10.1039/C5CC02626A

    31. [31]

      Cui W, Liu Q, Xing Z C, et al. Appl. Catal. B, 2015, 164:144-150 doi: 10.1016/j.apcatb.2014.09.016

    32. [32]

      Pan Y, Liu Y R, Zhao J C, et al. J. Mater. Chem. A, 2015, 3:1656-1665 doi: 10.1039/C4TA04867A

    33. [33]

      Xing Z C, Liu Q, Asiri A M, et al. Adv. Mater., 2014, 26:5702-5707 doi: 10.1002/adma.201401692

    34. [34]

      Radhakrishnan S, Kim H Y, Kim B S. Sens. Actuators B, 2016, 233:93-99 doi: 10.1016/j.snb.2016.04.056

    35. [35]

      Kannan P, Maiyalagan T, Marsili E, et al. Nanoscale, 2016, 8:843-855 doi: 10.1039/C5NR06802A

    36. [36]

      Yu S J, Peng X, Cao G Z, et al. Electrochim. Acta, 2012, 76:512-517 doi: 10.1016/j.electacta.2012.05.079

    37. [37]

      Wu W Q, Li Y B, Jin J Y, et al. Sens. Actuators, B, 2016, 232:633-641 doi: 10.1016/j.snb.2016.04.006

    38. [38]

      Prasad R, Gorjizadeh N, Rajarao R, et al. RSC Adv., 2015, 5:44792-44799 doi: 10.1039/C5RA03720D

    39. [39]

      Soomro R A, Ibupoto Z H, Sirajuddin, et al. Sens. Actuators B, 2015, 209:966-974 doi: 10.1016/j.snb.2014.12.050

    40. [40]

      Subramanian P, Niedziolka-Jonsson J, Lesniewski A, et al. J. Mater. Chem. A, 2014, 2:5525-5533 doi: 10.1039/C4TA00123K

    41. [41]

      Kannan P K, Rout C S. Chem. Eur. J., 2015, 21:9355-9359 doi: 10.1002/chem.v21.26

    42. [42]

      Guo Q H, Zhang M, Liu S W, et al. Anal. Methods, 2016, 8:8227-8233 doi: 10.1039/C6AY02299E

    43. [43]

      Yang J, Cho M S, Lee Y K. Sens. Actuators B, 2016, 222:674-681 doi: 10.1016/j.snb.2015.08.119

    44. [44]

      Ibupoto Z H, Tahira A, Mallah A B, et al. Electroanalysis, 2017, 29:213-222 doi: 10.1002/elan.v29.1

  • Figure 1  (a) XRD patterns of the Ni12P5 NPs; (b) Survey XPS spectrum for Ni12P5 NPs; High-resolution XPS spectra of Ni2p (c) and P2p regions (d)

    Figure 2  Typical TEM images of Ni12P5 NPs with low magnification (a) and high magnification (b)

    Figure 3  (a) CVs of Ni12P5/GCE (1, 2) and bare GCE (3, 4) in 0.1 mol·L-1 NaOH with the presence (1, 3) and absence (2, 4) of 0.1 mmol·L-1 glucose at a scan rate of 50 mV·s-1; (b) Amperometric response of the Ni12P5/GCE (1) and bare GCE (2) at 0.6 V with successive adding glucose of the same concentration in 0.1 mol·L-1 NaOH solution

    Figure 4  (a) CV curves of Ni12P5/GCE in 0.1 mol·L-1 NaOH with 0.1 mmol·L-1 glucose at various scan rates; (b) Linear relationship of oxidation current density vs the square root of scan rate

    Figure 5  (a) Amperometric response of the Ni12P5/GCE recorded at 0.60 V in 0.1 mol·L-1 NaOH solution with consecutive addition of various concentrations of glucose; (b) Linear fitting curve between current density and glucose concentrations

    Figure 6  (a) Amperometric response of the Ni12P5/GCE with consecutive additions of 1 mmol·L-1 glucose and 0.1 mmol·L-1 DA, UA, AA, Fru in 0.1 mol·L-1 NaOH solution at 0.60 V; (b) Variation in the response current density of Ni12P5/GCE toward 0.1 mmol·L-1 glucose in 0.1 mol·L-1 NaOH for 30 days

    Table 1.  Comparison between Ni12P5/GCE and other Ni-based electrochemical sensors for the detection of glucose

    Electrode material Sensitivity/
    (mA·L·mol-1·cm-2)
    Linear range/
    (mmol·L-1)
    Detection limit/
    (μmol·L-1)
    Reference
    Ni7S6/GCE 271.8 0.005~3.7 0.15 [37]
    NiO-MWCNT/CPEa 6 527 0.001~14 19 [38]
    NiO HHs-Nafion/GCE 1 052.8 0.5~4.5 1.2 [39]
    Ni(OH)2/rGO 11 400 0.02~30 15 [40]
    NiS 7.43 0.005~0.045 0.32 [41]
    Ni(OH)2/NCFb 1 342.2 0.005~9.15 0.8 [42]
    Ni2P/Ti 2 289.5 0.001~6.0 0.4 [22]
    c-Ni(OH)2 HRc 1 569 0.002~3.8 0.6 [43]
    Ni-NPs/TiO2NTs 700.2 0.004~4.8 2 [36]
    Ni12P5/GCE 1 572 0.002~4.2 0.8 This work
    aCPE: carbon paste electrode; bNCF: nitrogen-doped carbon foam; cHR: hollow nanorod
    下载: 导出CSV

    Table 2.  Test results of glucose in human blood serum (n=3)

    Sample Concentration of glucose/(mmol·L-1) Recovery/%
    Hospital Our sensor Added Measured value
    1 4.15 4.23 5 9.27 100.8
    2 4.96 5.02 5 10.15 102.6
    3 4.36 4.45 5 9.38 98.6
    下载: 导出CSV